BIPAP


Bilevel Positive Airway Pressure is a non-invasive treatment method for patients with sleep apnea with a mask that supplies compressed breathing air through a mask to prevent potential collapses or blockages in the lungs and breathing tubes. Allows the patient to breathe himself.

GOLDSLEEP Bipap S device also has a BACKUP FREQUENCY system to protect the patient. Thanks to this system, if the patient does not breathe for any reason, the device monitors the patient for 10 seconds and starts breathing at the end of this period.


Technical Specifications

IPAP

3 - 30 mbar (up to 30.0 mbar in 0.1 mbar increments) Breathing.

EPAP

3 - 30 mbar (up to 30.0 mbar in 0.1 mbar increments) Exhaling.

RAMP TIME

OFF / 5-10-15…90 MIN

RAMP PRESSURE

It can be adjusted between 3 - 30 mbar. It gradually rises to the IPAP / EPAP value.

I-SLOPE

OFF / 1-2-3-4-5 Length/rate of patient's breathing time (Time taken by the device to go from EPAP (positive exhalation pressure level) to IPAP (positive exhalation pressure level).

E-SLOPE

OFF / 1-2-3-4-5 Length/rate of patient's exhalation time (Time taken by the device to descend from IPAP (positive inspiratory pressure level) to EPAP (positive exhalation pressure level).

I SENSITIVITY

OFF / 1-2-3-4-5 It is the sensitivity to detect inspiration (the sensitivity to detect the beginning of the breathing phase of breathing.).

E SENSITIVITY

OFF / 1-2-3-4-5 It is the sensitivity to detect expiration (the sensitivity to detect the beginning of the exhalation phase of breathing.).

BACKUP FREQUENCY(RESPIRATORY SAFETY TIME)

ON / OFF – It is an additional safety measure to allow the patient to breathe. For example, if the patient does not breathe for 10 seconds, the device records this as apnea and activates after apnea. It works as a backup T mode and makes the patient breathe, then the device switches back to S mode.

TUBE DIAMETER

12mm/15mm/22 mm

SIGH

 

OFF / 300-350-400-450…1000 In one breath, it sends some extra air to the patient, allowing him to take a deep breath. The device uses this feature if the patient encounters conditions such as Apnea/Hypoapnea/Central Apnea during therapy. This feature is not suitable for use in patients with COPD.

PRESSURE UNIT

(hPa/mbar/cmH2o)